SWISS optio pipeline isolator
The pipeline isolator can be used for the medical gases oxygen O2, nitrous oxide N2O, Air, vacuum, carbon dioxide CO2 and as well for the gases used to operate surgical instruments such as AIR 800 and nitrogen N2.
Depending on the version used, the pipeline isolator can be mounted in pipelines of Ø12, Ø15, Ø22 or Ø28 mm. By means of the solder ring fittings on both ends, the isolator is soldered into the corresponding pipeline on the intended electrical separation point.
The pipeline isolator has been reviewed and approved by ELECTROSUISSE according to the relevant guidelines and standards (IEC/EN).
The area of use of the pipeline isolator is the central gas supply of medical gases in hospitals and clinics.
CE marking according to the Guideline for Medical Products 93/42/EEC.
Download files
Technical data:
- Classification
equipment class according to CE guideline 93/42 EEC annex IX, class II a
- Pressure range
max. 8 bar (Air 800 or N2)
- Gas
O2, N2O, AIR, Vakuum, CO2, AIR 800, N2
- Inlet / Outlet
soldering fitting Ø12 mm, Ø15 mm, Ø22 mm, Ø28 mm
- Material
brass, isolator tube PC (polycarbonate), sealing material : EPDM
- Operating temperature
-20 to +60°C
- Dimensions
total length for all tube diameters: 182 mm (152 mm between soldering shoulders)
Ø12 mm, Ø15 mm: outer diameter Ø30 mm, on both sides hexagon SW30
Ø22 mm: outer diameter Ø 40 mm, hexagon SW36
Ø28 mm: outer diameter Ø 40 mm, hexagon SW45
- Weight
460 g (Ø12 mm), 770 g (Ø28 mm)